:MONItor:VOLTage:CONDition
Syntax
:MONItor:VOLTage:CONDition {<V|>V|NONE},{AND|OR|NONE}
:MONItor:VOLTage:CONDition?
Description
Set the voltage monitor condition of the monitor (the current channel).
Query the voltage monitor condition of the monitor (the current channel).
Parameter
Name
Type
Range
Default
{<V|>V|NONE}
Discrete
<V|>V|NONE
NONE
{AND|OR|NONE}
Discrete
AND|OR|NONE
NONE
Explanation
You can set the voltage monitor condition to "<V" (<voltage), ">V" (>voltage)
or "NONE". "NONE" means do not monitor the voltage.
The actual monitor condition is the logic combination of the voltage, current
and power.
monitor value.
Return Format
The query returns the voltage monitor condition and logic relation separated by
comma, for example, <V,AND.
Example
:MONI:VOLT:COND <V,AND /*Set the voltage monitor condition (the current
channel) to "<V,AND"*/
:MONI:VOLT:COND? /*Query the current voltage monitor condition (the
current channel) and the query returns
<V,AND*/
Related
Commands
:MONItor:CURRent:CONDition
:MONItor:CURRent[:VALue]
:MONItor:POWER:CONDition
:MONItor:POWER[:VALue]
:MONItor:VOLTage[:VALue]