:MONItor:VOLTage[:VALue]
Syntax
:MONItor:VOLTage[:VALue] {<value>|MINimum|MAXimum}
:MONItor:VOLTage[:VALue]? [MINimum|MAXimum]
Description
Set the voltage value of the monitor condition (the current channel).
Query the voltage value of the monitor condition (the current channel).
Parameter
Name
Type
Range
Default
<value>
Real
Refer to the "Explanation"
Explanation
For multi-channel models, the range of <value> is from 0 to the maximum voltage
of the current channel and the default is 0.5*rated voltage of the current channel;
for single-channel model, the range of <value> is from 0 to the maximum voltage
of the current range and the default is 0.5*rated voltage of range 1.
Return Format
The query returns the voltage of the current monitor condition (the current
channel), for example, 5.000.
Example
:MONI:VOLT 5 /*Set the voltage of the monitor condition (the current channel)
to 5V*/
:MONI:VOLT? /*Query the voltage of the current monitor condition (the
current channel) and the query returns 5.000*/
Related
Command