Programming Features

Sensor Enabled
When the sensor is enabled, and volume calibration is selected, a single
manual calibration is taken. It does not matter if multiple volumes are
programmed (such as when using stormwater or upset sampling). When the
sensor is enabled, a single “Manual Calibration” choice provides sufficient
information to adjust the sensor to all programmed volumes (see

Table 24

).

Calibration Procedure—Sensor Disabled:

Note

:

Before calibrating, make

certain the desired sample volume
in the Setup menu is selected.

1.

From the Main Menu, press

OPTIONS

.

2.

Press

VOLUME CALIBRATION

.

3.

Select the desired volume to calibrate if more than one is listed
(see

Table 24

).

4.

Slip the pump tube off the sample fitting (near where it exits the pump
housing) and direct the end of the pump tube into a graduated cylinder.

5.

Press

START PUMPING

. The pump will purge the line, and start pumping a

sample.

6.

Press any key when the desired sample volume has been reached in the
graduated cylinder. The pump will stop to allow the line to completely
drain and then reverse to post-purge the line.

7.

If the pumped sample volume is correct, press

DONE

. If not, repeat the

calibration cycle by pressing

TRY AGAIN

.

8.

When finished, re-install the pump tube onto the sample fitting.

Note: To check the sample size after calibration use the Manual Mode function key to

take a grab sample. Do not go back into calibration to check the volume since the
volume compensation is reset to zero at the start of a calibration.

If one or more Timed Rinses are selected (sensor disabled), then they must
be calibrated as well. The Timed Rinse will rinse the intake line up to 3 times
prior to taking a sample.

Table 24 Calibration Volumes

Liquid Sampler

Available Calibration Volumes

Enabled

Manual Calibration

Disabled

Main Sample Volume

Disabled/Stormwater

Main Sample Volume

First Flush Volume

Disabled/Upset Sampling

Main Sample Volume

Upset Sample Volume

Disabled/Storm & Upset

Main Sample Volume

Upset Sample Volume

First flush Volume

Disabled/If Timed Rinses Set

Timed Rinse (in addition to the above)