Submerged Pressure Sensor

8990sensors.fm

Section 4

4.3.2 Submerged Pressure Sensor Programming

1.

From the Main Menu, select

OPTIONS>LEVEL SENSOR.

2.

Select Submerged Xducer using the

CHANGE CHOICE

key and press

ACCEPT.

4.3.3 Submerged Pressure Sensor Calibration

In sites with harsh conditions (extremes of level, temperature, harsh
chemicals, etc.) calibration should be performed more often.

1.

From the Main Menu, select

OPTIONS > ADVANCED OPTIONS >

CALIBRATION > SUBMERGED PROBE.

2.

Choose the orientation that the sensor will be mounted in the flow stream,
horizontal or vertical, using the

CHANGE CHOICE

key. Press

ACCEPT

.

3.

Lift the sensor out of the water and hold it in the air in the same orientation
that you selected in the previous step (horizontal or vertical) (

Figure 18

).

Then press

ACCEPT

to continue.

Figure 18

Lifting the Sensor Out of the Water

4.

Follow either the vertical or horizontal procedure below.

Table 6 Submerged Level Sensor Base Board Connection (J21)

Pin

Signal Description

Wire Color

A

V +

Red

B

Out +

Yellow

C

Out -

Green

D

Ground

Black

11:00 AM 21 - APR - 01

CALIBRATION

ACCEPT

ORIENTATION OF

CHANGE

CHOICE

SUBMERGED PROBE:

CANCEL

HORIZONTAL

SELECT APPROPRIATE UNITS

Horizontal

Vertical