Basic Programming Setup

8990startup.fm

Section 3

a.

In the Sample Collection menu, press

CHANGE CHOICE

until Flow

Proportional is displayed. Press

ACCEPT

.

b. In the Flow Pacing menu, press

CHANGE CHOICE

until

Const Time/Var Vol appears. Press

ACCEPT

.

c.

In the Average Flow Rate menu use the numeric keypad to enter the
known historical Average Flow Rate for a particular site.

d. In the Interval menu use the numeric keypad to enter the time interval

between sample collections. Press

ACCEPT

.

e.

In the Total Volume Desired menu use the numeric keypad to enter
the total volume of the sample. Press

ACCEPT

.

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

ACCEPT

CHANGE

UNITS

SAMPLE COLLECTION:

CANCEL

FLOW-PROPORTIONAL

ENTER: 1.00 - 99999999

11:00 AM 21 - APR - 01

FLOW PACING

ACCEPT

CHANGE

CHOICE

FLOW PACING MODE:

CANCEL

CONST TIME/ VAR VOL

CHOICES: VAR T / CST VOL, CST T / VAR VOL

11:00 AM 21 - APR - 01

CONST TIME/ VAR VOL

ACCEPT

AVERAGE FLOW RATE:

150.00 gph

CANCEL

CLEAR

ENTRY

ENTER: 1.00—99999999

11:00 AM 21 - APR - 01

CONST TIME/ VAR VOL

ACCEPT

INTERVAL:

0:02 (hrs:min)

CANCEL

CLEAR

ENTRY

ENTER: 1.00—999:00 (hrs:min)

11:00 AM 21 - APR - 01

CONST TIME/ VAR VOL

ACCEPT

TOTAL VOLUME DESIRED:

1500 Ml

CANCEL

CLEAR

ENTRY

ENTER: 10—9999