Hach 900 MAX User Manual
Page 46

Page 44
Basic Programming Setup
8990startup.fm
Section 3
a.
In the Sample Collection menu, press
CHANGE CHOICE
until Flow
Proportional is displayed. Press
ACCEPT
.
b. In the Flow Pacing menu, press
CHANGE CHOICE
until
Const Time/Var Vol appears. Press
ACCEPT
.
c.
In the Average Flow Rate menu use the numeric keypad to enter the
known historical Average Flow Rate for a particular site.
d. In the Interval menu use the numeric keypad to enter the time interval
between sample collections. Press
ACCEPT
.
e.
In the Total Volume Desired menu use the numeric keypad to enter
the total volume of the sample. Press
ACCEPT
.
11:00 AM 21 - APR - 01
SAMPLE COLLECTION
ACCEPT
CHANGE
UNITS
SAMPLE COLLECTION:
CANCEL
FLOW-PROPORTIONAL
ENTER: 1.00 - 99999999
11:00 AM 21 - APR - 01
FLOW PACING
ACCEPT
CHANGE
CHOICE
FLOW PACING MODE:
CANCEL
CONST TIME/ VAR VOL
CHOICES: VAR T / CST VOL, CST T / VAR VOL
11:00 AM 21 - APR - 01
CONST TIME/ VAR VOL
ACCEPT
AVERAGE FLOW RATE:
150.00 gph
CANCEL
CLEAR
ENTRY
ENTER: 1.00—99999999
11:00 AM 21 - APR - 01
CONST TIME/ VAR VOL
ACCEPT
INTERVAL:
0:02 (hrs:min)
CANCEL
CLEAR
ENTRY
ENTER: 1.00—999:00 (hrs:min)
11:00 AM 21 - APR - 01
CONST TIME/ VAR VOL
ACCEPT
TOTAL VOLUME DESIRED:
1500 Ml
CANCEL
CLEAR
ENTRY
ENTER: 10—9999