Specifications

8990spc.fm

Specifications

Rain Gauge Input

General Information

For use with the Tipping Bucket Rain Gauge.

The Sampler Program can be initiated upon field selectable rate of rain.

Sampler records rainfall data.

Each tip = 0.25 mm (0.01in.) of rain.

Analog Input Channels

General Information

Up to 3 additional data logging channels record data from external source(s)

Field assignable units

-4.0 to +4.0 V dc and 0 to 20 mA

4–20 mA Output

General Information

Up to 2 output signals available.

User assignable

Maximum Resistive
Load

600 ohms

Output Voltage

24 V dc–no load

Insulation Voltage

Between flow meter and 4–20 mA output—2500 V ac

Between the two 4–20 mA outputs—1500 V ac

Alarm Relays

General Information

(4) 10 amp/120 V ac or 5 amp/220 V ac form C relays

±0.1% FS Error

User assignable for any internal or external data channel or event.

Downlook Ultrasonic Sensor 40 kHz

Accuracy

0.003 m (±0.01 ft)

Maximum Range

3.35 m (11 ft) with a 3.05 m (10 ft) span

Deadband

38 cm (15 in.) maximum, self-minimizing

Material

PVC housing with Buna-N acoustic window

Cable

4 conductor with integral stainless steel support cable

Submerged Pressure Transducer

Material

Epoxy body with stainless steel diaphragm.

Cable

Polyurethane sensor cable with air vent.

Cable Length

7.6 m (25 ft) standard; 76 m (250 ft) maximum

Sensor Dimensions

2 cm H x 3.8 cm W x 12.7 cm L (0.8 in. x 1.5 in. x 5 in.)

Maximum Range

5 psi, 0.063 to 3.5 mm (0.018 to 11.5 ft)

15 psi, 0.063 m to 10.5 m (0.018 to 34.6 ft)

Maximum Allowable
Level

3x over pressure

Operating Temperature
Range

0 to 71 °C (32 to 160 °F)

Compensated
Temperature Range

0 to 30 °C (32 to 86 °F)

Air Intake

Atmospheric pressure reference is desiccant protected.