Section 6

Page 87

8990cm.fm

4–20 mA Option

6.3 4–20 mA Option

The 4–20 mA option provides a current loop for controlling external devices
such as a chart recorder or PC. Either one or both of the 4–20 mA outputs can
be factory installed and are isolated from each other.

6.3.1 4–20 mA Connection

The interface has a 3-ft cable with a connector on one end, and a 10-ft cable
with two open wire leads on the other. Insert the connector into the sampler
receptacle labeled “Auxiliary”, located on the left side of the control housing.
On the 10-ft cable, the wire with clear insulation is positive (+) and the wire
with black insulation is negative (-).

Rating:

Isolation Voltage:
Between sampler and either 4–20 mA output: 2500 V ac
Between the two 4–20 mA outputs: 1500 V ac

Maximum Resistive Load: 600 ohm

Output Voltage: 24 V dc - no load

Cable Required
4–20 mA Interface Cable (Cat. No. 2924), 25 ft long, 4-pin connector on one
end, tinned wire leads on the other end.

6.3.2 4–20 mA Programming

1.

From the Main Menu, select

OPTIONS > ADVANCED OPTIONS > 4–20 mA

OUTPUTS > SELECT

.

Note: When the 4–20 mA outputs
are disabled and not completely
turned off, they will continue to
output a steady 4 mA.

2.

Enable the 4–20 mA outputs by pressing

CHANGE CHOICE

while in the

4–20 mA output menu.

3.

When the display shows the outputs as enabled, press

ACCEPT

.

4.

Choose either OUTPUT A or OUTPUT B. Use the

UP

and

DOWN

keys to

highlight the choice, then press

SELECT.

Table 16 4–20 mA Connections (J18)

Pin

Signal Description

Wire Color

A

Output 1 + (pos)

Yellow

B

Output 1 - (neg)

Black

C

Output 2 + (pos)

Red

D

Output 2 - (neg)

Green

11:00 AM 21 - APR - 01

4–20 mA OUTPUTS

SELECT

OUTPUT A

OUTPUT B

RETURN

Advertising