Hach 900 MAX User Manual
Page 88

Page 86
Modem
8990cm.fm
Section 6
Table 15 Pager Alarm Codes
Alarm
Code
Equipment
Reason
Alarm
Code
Equipment
Reason
Low Main Battery
1
—
Battery pack is less than 11.5 V
High CH5
28
—
—
Memory Battery
2
—
Internal memory battery is low
High CH6
29
—
—
Low Slate Memory
3
—
Less than 10% slate memory left
High CH7
30
—
—
Slate Memory Full
4
—
Slate memory is used up
High
Reference
Temp.
31
—
—
Modem Failure
5
—
Modem chip/modem board failure
High
Velocity
32
—
—
Missed Sample
6
—
No liquid detected when sampling
High D.O.
33
—
—
Purge Failure
7
—
Water present at sensors after
purge
High D.O.
Temp.
34
—
—
Jammed Distributor
8
—
Sensors indicate arm is not
moving
High
Conductivity
35
—
—
Bottle is Full
9
—
Full bottle option indicator is on
High Cond.
Temp.
36
—
—
U-Sonic Echo Loss
10
—
No return signal detected
Low Level
37
—
—
Xducer Ringing
11
—
The return signal is detected too
soon
Low Flow
38
—
—
U-Sonic failure
12
—
Ultrasonic board detects an error
Low
pH/ORP
39
—
—
RS485 Timed Out
13
—
Comm. problems with RS485
Low
Process
Temp.
40
—
—
Unable to Cool
14
AWRS only
Too high temperature in a
compartment
Low CH1
41
—
—
Unable to heat
15
AWRS only
Too cold a temperature in
compartment
Low CH2
42
—
—
Low Bubbler Pres.
16
(not
applicable)
Possible leak in bubble tank
Low CH3
43
—
—
Clogged Bubbler
17
(not
applicable)
Bubbler tube is plugged
Low CH4
44
—
—
High Level
18
—
—
Low CH5
45
—
—
High Flow
19
—
—
Low CH6
46
—
—
High Flow Rate of
Chg.
20
—
—
Low CH7
47
—
—
High pH/ORP
21
—
—
Low
Reference
Temp.
48
—
—
High Process
Temperature
22
—
—
Low Velocity
49
—
—
High Rainfall
23
—
—
Low D.O.
50
—
—
High CH1
24
—
—
Low D.O.
Temp.
51
—
—
High CH2
25
—
—
Low
Conductivity
52
—
—
High CH3
26
—
—
Low Cond.
Temp.
53
—
—
High CH4
27
—
—