Section 6 communication setup, 1 rs232 cable, 1 rs232 connection – Hach 900 MAX User Manual
Page 81: 2 rs232 programming

Page 79
8990cm.fm
Communication Setup
Section 6
Communication Setup
Communication setup details connections and programming features of the:
6.1 RS232 Cable
6.1.1 RS232 Connection
The RS232 connector is a serial input/output port for communicating with a
sampler from an external device such as a Data Transfer Unit (DTU-II) or
direct serial connection to a personal computer with application software.
Cable Required
RS232 Sampler/Flow Meter to PC Cable Assembly (Cat. No. 1727), 10 ft long,
6-pin connector on one end, 9-pin D connector on the other end (9-pin to
25-pin D type adapter included).
6.1.2 RS232 Programming
Note: Some early generation IBM
compatibles may have a serial port
that is not capable of
communicating reliable at 19,200
baud. If errors develop at high baud
rates, try lowering the baud rate one
step at a time (on both the sampler
and Streamlogg II™, InSight
®
, or
Vision
®
software) until error free
communications is achieved.
1.
From the Main Menu, select
OPTIONS > ADVANCED OPTIONS >
COMMUNICATIONS SETUP > RS232 SETUP
.
2.
Select a baud rate for data communications by pressing the
CHANGE
CHOICE
key to cycle through the possible choices; 1200, 2400, 4800,
9600 or 19,200 baud. The higher the baud rate setting, the faster data will
transfer. Set the baud rate to the highest setting allowed by the computer.
The baud rate must correspond to the baud rate selected in the software.
Press
ACCEPT
.
Note: Long runs of RS232 cable, especially if they are run near large motors or
fluorescent lights can cause communication errors and may require a slower
baud rate.
•
RS232 Serial Port (
•
Optional Modem (
)
•
4–20 mA Option (
•
Alarm Relays (
)
•
Analog Inputs (
)
Table 13 RS232 Connection
Pin
Signal Description
Wire Color
A
not used
White
B
ground
Blue
C
DSR
Yellow
D
RCD
Black
E
DTR
Red
F
TXD
Green