Fluke 434 User Manual

Page 106

Advertising
background image

Fluke 433/434

Users Manual

21-4

Measurement Measurement

Range Resolution

Accuracy

Relative Amplitude

0.0 ... 100.0 %

0.1 %

± 0.1 % ± n x 0.1 %,
(±0.4% for %r)

Absolute Amplitude

Vnom: 60 ... 125 Vrms

Vnom: 125 ... 250 Vrms

Vnom: 250 ... 500 Vrms

0.0 ... 250.0 Vrms

0.0 ... 500.0 Vrms

0.0 ... 999.9 Vrms

0.1 Vrms

± 5 % of meas ± 2
counts

Phase

-360º ... +360º

± n × 1.5º

Frequency

0 ... 3500 Hz

1 Hz

± 1Hz

THD

0.0 ... 100.0 %

0.1 %

± 2.5 %

DC relative

absolute

0.0 ... 100.0 %

0.0 ... 100.0 V

0.1 %

0.1 V

± 1 %

± 5 % of meas ± 10
counts


Voltage Dips

Settings Adjustment

Range

Resolution

Additional

Information

Dip Threshold level

50.0 ... 100.0 % of Vnom 0.1 %

Dip Hysteresis level

0.0 ... 10.0 % of Vnom

0.1 %

results based on ½ cycle
rms

Measurements Measurement

Range Resolution

Accuracy

Dip Magnitude

0.0 % ... 100.0 % of
Vnom

0.1 %

± 1 % of Vnom

Dip Duration

hhh,mm,ss,mmm

10 ms

± 20 ms (at F=50 Hz)


Voltage Swells

Settings Adjustment

Range

Resolution

Additional

Information

Swell Threshold level

100.0 ... 200.0 % of
Vnom

0.1 %

Swell Hysteresis level

0.0 ... 10.0 % of Vnom

0.1 %

results based on ½ cycle
rms

Measurements Measurement

Range Resolution

Accuracy

Swell Magnitude

100.0 %...200.0 % of
Vnom

0.1%

± 1 % of Vnom

Swell Duration

hhh,mm,ss,mmm

10 ms

± 20 ms (at F= 50 Hz)