The Bar Graph display shows the percentage contribution of each of the components

related to the full signal. A signal without distortion should show a 1

st

harmonic (= the

fundamental) at 100 % while the others are at zero: in practice this will not occur because

there always is a certain amount of distortion resulting in higher harmonics.
A pure sinewave becomes distorted when higher frequency components are added to it.

Distortion is represented by the THD percentage. The display can also show the

percentage of the DC component and the K-factor. The K-factor is a number that

quantifies potential losses in transformers due to harmonic currents. Higher order

harmonics influence the K-factor more than low order harmonics.
The table below shows the number of Bar Graphs displayed simultaneously in one

screen:

Harmonics

Harmonics & Interharmonics

All phase display

1 ... 12

1 ... 6

Single phase display

1 ... 50

1 ... 25

The left/right arrow keys are used to position the Cursor on a particular bar. The screen

header will show for that bar phase identifier, harmonic number, frequency, and phase

angle. If not all bars are shown on the screen, you can bring the next set within the

viewing area by moving the Cursor off the left or right end of the screen. The up/down

arrow keys are used for vertical zoom: 100 %, 50 %, 20 %, 10 %, or 5 % at full scale can

be selected. Using the SETUP key and function key F3 - FUNCTION PREF you can

choose harmonics display as a percentage of the fundamental voltage (%f) or the total of

harmonic voltages (%r). For detailed information see Chapter 18, FUNCTION

PREFerences.
Available function keys:

F1

Selection of harmonics type: Voltage, Current, or Real
Power (Watt). Power harmonics can have positive and
negative polarity.

F2

Selection of waveform set to be used: A (L1), B (L2), C
(L3), N (neutral) or ALL

F3

Access the table screen.

F4

Interharmonics display on/off (Fluke 434 only).

F5

Switch between HOLD and RUN of screen update.
Switching from HOLD to RUN invokes a menu to select
immediate (NOW) or TIMED start time which allows you
to define start and duration of the measurement.