Trend recording – Fluke 434 User Manual

Page 113

Advertising
background image

Specifications

Trend Recording

21

21-11

Energy
Whr, VAhr, VARhr Ranges:

V

1 V10 V100 V1000

0.1 mV/A

000.0 kWhr ...

200.0 GWhr

0.000 MWhr ...

2.000 TWhr

00.00 MWhr ...

20.00 TWhr

000.0 MWhr ...

200.0 TWhr

1 mV/A

00.00 kWhr ...

20.00 GWhr

000.0 kWhr ...

200.0 GWhr

0.000 MWhr ...

2.000 TWhr

00.00 MWhr ...

20.00 TWhr

10 mV/A

0.000 kWhr ...

2.000 GWhr

00.00 kWhr ...

20.00 GWhr

000.0 kWhr ...

200.0 GWhr

0.000 MWhr ...

2.000 TWhr

100 mV/A

000.0 Whr ...

200.0 MWhr

0.000 kWhr ...

2.000 GWhr

00.00 kWhr ...

20.00 GWhr

000.0 kWhr ...

200.0 GWhr

1 V/A

00.00 Whr ...

200.0 kWhr

000.0 Whr ...

200.0 MWhr

0.000 kWhr ...

2.000 GWhr

00.00 kWhr ...

20.00 GWhr

Maximum Integration Time: 9999 hours


Whr, VAHr, Resolution and Accuracy:

Maximum Resolution (lowest range)

Accuracy

(excl. Clamp

error)

V

1 V10 V100 V1000

0.1 mV/A

100 Whr

1 kWhr

10 kWhr

100 kWhr

1 mV/A

10 Whr

100 Whr

1 kWhr

10 kWhr

10 mV/A

1 Whr

10 Whr

100 Whr

1 kWhr

100 mV/A

0.1 Whr

1 Whr

10 Whr

100 Whr

1 V/A

0.01 Whr

0.1 Whr

1 Whr

10 Whr

± 1.5% of meas
± 10 counts

Note: The Usage scale (in Whrs) starts a factor 10 lower than the equivalent Power scale

(in W). This means that after 6 minutes the Usage number and the Power number are of

the same magnitude.

Trend Recording

General

Item Specification

Resolution

1s, 5s, 30s, 1m, 5m, 15m, 30m, 1h, 3h, 6h

Duration

0.5h, 2.5h, 7.5h, 15h, 30h, 150h, 450h, 900h, 75d, 225d, 450d

Memory

1800 min, max, and avg. points for each reading

Advertising
This manual is related to the following products:

433