Fluke 433/434

Users Manual

21-8

Inrush Current

Settings Adjustment

Range

Resolution

Additional

Information

Inrush Threshold level

0 ... 999 % of Inom

1 %

Inrush Hysteresis level

0 ... 999 % of Inom

1 %

Results based on Irms

1/2

( Itrh - Ihys > Inom )

Inrush Evaluation time

7.5 s, 15 s, 30 s, 1.5 m,
3 m, 6 m, 12 m, 30 m

fixed
ranges

Measurements

Measurement Range

Resolution

Accuracy ( excl.

Clamp error )

Inrush Magnitude

0.1 mV/A

1 mV/A

10 mV/A

100 mV/A

1 V/A

0.00 ... 20.00 kArms

0 ... 2000 Arms

0.0 ... 200.0 Arms

0.00 ... 20.00 Arms

0.000 ... 2.000 Arms

10 Arms

1 Arms

0.1 Arms

0.01 Arms

0.001 Arms

± 1 % of meas ± 5
counts

Inrush Duration

mm:ss:mmm

10 ms

± 20 ms (at F = 50 Hz)

Current Magnitude

0.1 mV/A

1 mV/A

10 mV/A

100 mV/A

1 V/A

0.00 ... 20.00 kArms

0 ... 2000 Arms

0.0 ... 200.0 Arms

0.00 ... 20.00 Arms

0.000 ... 2.000 Arms

10 Arms

1 Arms

0.1 Arms

0.01 Arms

0.001 Arms

± 1 % of meas ± 5
counts


Current Unbalance

Measurements

Measurement Range

Resolution

Accuracy ( excl.

Clamp error )

Negative Unbalance
Ratio

0.0 ... 20.0%

0.1%

± 1%

Zero Unbalance Ratio

0.0 ... 20.0%

0.1%

± 1%