Electro magnetic compatibility (emc), Safety – Fluke 434 User Manual

Page 118

Advertising
background image

Fluke 433/434

Users Manual

21-16

Electro Magnetic Compatibility (EMC)

Item

Customer Specification

Additional Information

Emission and Immunity

EN-61326

Fluke 433/434, including standard
accessories, conforms with the
EEC directive 89/336 for EMC
immunity, as defined by EN-
61326, with the addition of the
table below

Frequency

Disturbance < 0.5 %

Disturbance < 10 %

80 – 400 MHz

All ranges

400 – 600 MHz

All other ranges

125 V range

600 MHz – 1 GHz

All ranges

The Analyzer is susceptible for RF fields with a field strength of 10 V/m, between 400 and 600 MHz
(Performance criteria B).

Safety

Item

Customer Specification

Additional Information

including approval
Reference Standards

EN/IEC61010-1 2nd edition
1000V Measurement Category III,
600V Measurement Category IV,
Pollution Degree 2.
ANSI/ISA S82.01-1994
CAN/CSA C22.2 No. 61010-1-04
(including approval)

According to CE marking

Max voltage between any
Voltage banana input and
safety ground

1000 V CAT III 600 V CAT IV

At altitude 2000 m … 3000 m:
1000 V CAT II, 600 V CAT III,
300 V CAT IV

42 Vpeak Max voltage on
Current BNC input

REMARK: the BNC grounds of
the BNC’s are connected to the
ground banana input

Advertising
This manual is related to the following products:

433