READY STATUS REGISTER

The RPM4 contains an 8 bit Ready Status Register that reflects specific RPM4 Q-RPT
measurement ready events. Enabled events in this register will set or clear the RSB bit of the
Status Byte Register.

Table 18. 8 bit ready status register

N/A

(128)

MEAS

LO

(64)

NRDY

LO

(32)

RDY

LO

(16)

N/A

(8)

MEAS

HI

(4)

NRDY

HI

(2)

RDY

HI

(1)

This register can be read using the “*RSR?” query, Each of these status bits can set the
RSB bit of the Status Byte Register, causing a SRQ to occur IF the RSB bit is enabled to do so.
The Standard Event Status Enable Register (“*RSE” program message ) determines which
of these flags are able to assert the RSB bit. The description of these bits are given as:

MEAS LO Lo-RPT measurement ready Bit 6 (64)

Indicates that the Hi Q-RPT has completed a Q-RPT measurement.

NRDY LO Lo Q-RPT pressure Not Ready Bit 5 (32)

Indicates that the Hi Q-RPT made a transition from Ready to Not Ready as
defined by the stability settings (see Section 3.2.1).

RDY LO

Lo Q-RPT pressure Ready Bit 4 (16)

Indicates that the Hi Q-RPT is Ready as defined by the stability settings (see
Section 3.2.1).

MEAS HI

Hi Q-RPT measurement ready Bit 2 (4)

Indicates that the Hi Q-RPT has completed a Q-RPT measurement.

NRDY HI

Hi Q-RPT pressure Not Ready Bit 1 (2)

Indicates that the Hi Q-RPT made a transition from Ready to Not Ready as
defined by the stability settings (see Section 3.2.1).

RDY HI

Hi Q-RPT pressure Ready Bit 0 (1)

Indicates that the Hi Q-RPT is Ready as defined by the stability settings (see
Section 3.2.1).