RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL

© 2007 DH Instruments, a Fluke Company

Page 42

Run AutoZ by COM2

AutoZ by COM2 allows the value of P

std,0

(see Section 3.5.1

 Principle) to be read

automatically from a DHI RPMx connected to the RPM4 COM2 communications
port. The RPMx is read and the new P

offset

is calculated automatically.

To access run AutoZ by COM2 press [AutoZ], <2COM2>. The display is:

1. Real time pressure reading (without head

correction), pressure unit of measure and
measurement mode of the active Q-RPT.

2.

Active Q-RPT position designator.

3.

Real time reading from the RPMx connected
to RPM4’s COM2 communications port.

14.03 psi a Hi

Pstd,0:0.00

Observe the displayed pressures and verify that they are stable. When ready,
press [ENT] to cause AutoZ to run. RPM4 logs both readings and calculates a
new AutoZ offset value. The display is:

Press [ENT] to activate the new value of
P

offset

or [ESC] to start over.

Old P

offset

: 0.0 Pa

New P

offset

: 2745 Pa

For RPM4 to communicate with an RPMx connected to its COM2 port, the

RPM4 and the RPMx RS-232 interfaces must be set up properly

(see Section 3.5.2). If, the RPM4 is unable to communicate with an RPMx via COM2

when running AutoZ by COM2, it times out after 6 seconds and displays <RPM NOT

detected>.

The value of P

offset

is always displayed and entered in Pascal (Pa).

Run AutoZ by Lo RPT

AutoZ by Lo RPT is available only when the active Q-RPT is the Hi Q-RPT.
AutoZ by Lo RPT allows the Lo Q-RPT in an the RPM4 to serve as the AutoZ
reference (source of P

std,0

) (see Section 3.5.1,

 PRINCIPLE).

The Lo Q-RPT is read and the new P

offset

is calculated automatically.

To access run AutoZ by Lo Q-RPT press [AutoZ], <3Lo RPT >. The display is:

1. Real time pressure reading (without head

correction), pressure unit of measure and
measurement mode of the active Q-RPT.

2.

Active Q-RPT position designator.

3.

Real time reading from the Lo Q-RPT.

14.03 psi a Hi

Pstd,0: 14.08

Observe the displayed pressures and verify that they are stable. When ready,
press [ENT] to cause AutoZ to run. RPM4 logs both readings and calculates a
new AutoZ offset value. The display is:

Press [ENT] to activate the new value of
P

offset

or [ESC] to start over.

Old P

offset

: 0 Pa

New P

offset

: -345 Pa

The value of P

offset

is always displayed and entered in Pascal (Pa).