Fluke HPMS User Manual

Page 94

Advertising
background image

RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL

© 2007 DH Instruments, a Fluke Company

Page 84

CONTn
Purpose

Have the RPM4 reply continuously as each new measurement becomes available.

Query “CONTn?”
Classic “CONTn
Optional Suffix

“n”

The active Q-RPT is assumed if no suffix is given.
‘1’ Specify the Hi Q-RPT
‘2’ Specify the Lo Q-RPT

Remarks

This command is somewhat different from most as it enables the RPM4 to continuously output the
selected Q-RPT’s measurement without further queries. This continues until the next query is
received by the RPM4, or if the ABORT” or “*CLS” command is received. The frequency of the
measurements is determined by the readrate of the Q-RPT. The data returned includes the
pressure of the specified or active Q-RPT, current unit of measure and measurement mode.

Example
(enhanced)

Query sent:
Query reply:

“CONT1?”
“1936.72 kPa a” (repeats every measurement cycle)

Example
(classic)

Query sent:
Reply:

“CONT”
“1936.72 kPa a” (repeats every measurement cycle)

See Also

“PR”, “READRATE”, “ABORT”

DATE
Purpose

Read or set the RPM4 date.

Command
Query

“DATE date
“DATE?”

Classic “DATE= date

“DATE”

Arguments

date:

The date in the numerical only format “YYYYMMDD”

Remarks

The RPM4 has an internal real time calendar clock. The span of acceptable dates ranges
from 19800101 to 20791231. The reply is always in the YYYYMMDD format.

Example
(enhanced)

Cmd sent:
Query reply:

“DATE 20030115”
“20030105”

(No reply if GPIB-488)

Example
(enhanced)

Cmd sent:
Query reply:

“DATE? 20030105”
“20030105”

Example
(classic)

Sent:
Reply:

“DATE=20021201”
“20021201”

Errors

ERR# 6:

Missing or improper program message argument(s).

See Also

“TIME”
3.5.5.3

ERR
Purpose

Read the new available error message from the Error Queue.

Query “ERR?”
Classic “ERR”
Remarks

This program message obtains additional details about an error that has occurred. If the user
receives an “ERR# nn” reply, or the enhanced mode is enabled using the IEEE-488 interface and
an error has been detected, the error is put into a FIFO Error Queue. The “ERR” program
message pulls and replies the oldest error message available. In “classic” mode only the most recent
error can reside in the queue. “OK” is replied if there are no error messages left. If an error occurs in
enhanced mode using the IEEE-488 interface, no reply is generated even if the command was a
query.

Example
(enhanced):

Query sent:
Query reply:

“ERR?”
“Numeric argument missing or out of range”

Example
(classic)

Sent:
Reply:

“ERR”
“Numeric argument missing or out of range”

See Also

4.4.2