2 deleting autorange ranges, 2 <2res> (resolution), Deleting autorange ranges – Fluke HPMS User Manual

Page 58: 2res> (resolution), 2 deleting a, Ranges

Advertising
background image

RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL

© 2007 DH Instruments, a Fluke Company

Page 48

Press [ENT] to save the range or [ESC] to return to the MAIN RUN screen without
saving the range.

Default (<DF>) Q-RPT ranges cannot be saved. They are permanent.

Duplicate AutoRange ranges cannot be saved. A duplicate AutoRange range is

a range that has the same measurement mode and full scale and uses the same Q-
RPT as a range that has already been saved.

3.4.1.2

DELETING A

UTO

R

ANGE

RANGES

 OPERATION

To delete a single saved AutoRange range, press [SETUP], <1range>, <2delete>.
If there are any saved ranges available to delete, the display is:

1.

Indication that the cursor control keys may be
used to toggle through the saved ranges to
select the one to be deleted.

2.

Designator and position of the Q-RPT used by
the range to be deleted.

3.

Pressure unit of measure of the range to be
deleted.

4.

Full scale pressure of the range to be deleted
in gauge mode and absolute mode if available.

< and > A70M Hi

psi 4000g/4000a

Use the cursor control keys to toggle through the ranges available to be deleted.
When the desired range to be deleted is displayed, press [ENT]. If you are sure you
want to delete this range, press [ENT] again in response to the <Delete?> query.
[ESC] returns to the <AutoRange:> screen without deleting a range.

To delete all of the saved AutoRange ranges, press [SETUP], <1range>,
<3delete all>. If you are sure you want to delete all the ranges, respond <1yes>
to the <Delete all saved ranges?> query. If not, respond <2no>.

3.4.2

<2RES> (RESOLUTION)

 PURPOSE

To set the resolution with which measured pressures and other pressure indications and
settings are displayed.

 PRINCIPLE

The resolution with which RPM4 displays pressure values can be adjusted. This feature can be
used to reduce the resolution when lower precision measurements are being made and excess
resolution might confuse or distract the operator.

The resolution setting determines the number of digits with which pressure is displayed. The
resolution is calculated based on the span of the range and then rounded to the furthest digit to
the right. For example, resolution of 0.001 % on a range of 1 500 psi is 1 500 X 0.001 % = 0.015
which is rounded down to 0.01 psi.

Default resolution is 10 ppm of active range span. Resolution is set automatically by

AutoRange (see Section 3.3.4) and by AutoTest initialization (see Section 3.3.10).

Advertising
This manual is related to the following products: