© 2007 DH Instruments, a Fluke Company

Recommendations for the Use of the AutoZ Function

The AutoZ function provides a powerful and easy to use tool for improving the stability over
time of RPM4 Q-RPTs and maximizing the recalibration interval by compensating for zero
drift between full recalibrations. The following simple recommendations will help assure that
you use this feature to best advantage.

In gauge mode: Always leave AutoZ ON. Use [AutoZ] to run AutoZ whenever RPM4 is

vented and displays a significantly non-zero value. Wait at least 2 minutes after venting before
running AutoZ.

In absolute measurement mode: Always leave AutoZ ON. Use [AutoZ] to run AutoZ every

approximately every 30 days or when RPM4 has been exposed to a temperature change
exceeding 15 ºC (36 ºF). AutoZero the Lo Q-RPT against a barometer with measurement
uncertainty of ± 0.04 psi (275 Pa) or better (note that this is only 0.3% of atmospheric pressure
so a relatively low accuracy barometer is adequate). AutoZero the Hi Q-RPT against the Lo Q-
RPT. Wait 5 to 10 minutes before running AutoZero in absolute mode.

 OPERATION

The AutoZ function and values are Q-RPT AND measurement mode (gauge or absolute)

specific.

To RUN the AutoZ routine that rezeroes the Q-RPT, use [AutoZ] (see Section 3.3.9).

To access the RPM4 AutoZ function press [SPECIAL], <1AutoZ>. The display is:

1.

Active Q-RPT designator.

2.

Indication of active measurement mode (<abs> for absolute,
<gage> for gauge).

3.

Indication of whether AutoZ is currently ON or OFF for this
Q-RPT and measurement mode.

1off 2view Hi

3edit abs ON

Select <1off> (or <1on>) to change the AutoZ status for the current Q-RPT and
measurement mode from ON to OFF or vice versa.

AutoZ ON is indicated by a <z> in the MAIN RUN screen, top line, third character from the

right. When AutoZ is OFF, the character is blank.

Select <2view> to view the current value of P

offset

for the active Q-RPT and measurement mode.

P

offset

should be zero in absolute measurement mode when the RPM4 is new or has just been

calibrated. P

offset

should be roughly equal to atmospheric pressure when the Q-RPT is in gauge

measurement mode.

Select <3edit> to edit the value of P

offset

for the active Q-RPT and measurement mode

(see Section 3.5.1.1).

The value of P

offset

is always displayed and entered in Pascal (Pa).