Fluke HPMS User Manual

Page 63

Advertising
background image

3. OPERATION

Page 53

© 2007 DH Instruments, a Fluke Company

A File AutoTest definition file defines:

• The identifying number of the File AutoTest definition file.
• Pressure measurement mode (gauge or absolute).
• Pressure unit of measure.
• DUT full scale in pressure unit of measure and measurement mode.
• DUT tolerance (% FS).
• Number of test points or pressure targets to include in the AutoTest.
• Numerical value of each test point or pressure target.

See Section 3.3.10,

 OPERATION for information on running a Quick or File

AutoTest.

See Section 3.4.5.1 for information on viewing the data files that result from
running a Sequence.

 OPERATION

See Section 3.3.10, OPERATION, for information on RUNNING a Quick or File

AutoTest.

To access the AutoTest definition viewing and editing functions press [SETUP],
<5ATest>, <2file>. Select <1view> to view an AutoTest definition file. Select
<2edit> to edit an AutoTest definition file.

When <2edit> is selected the display is:

Edit File AutoTest #

1

Enter the number (1 - 40) of the File AutoTest to be edited.

To create a File AutoTest definition file similar to an existing definition file

without entering all the definition information, edit an existing definition file and at

the end of the editing process, save it as a different file number.

The display is:

Measurement mode:

1absolute 2gauge

The cursor is on the current selection. Select the desired pressure measurement
mode for the AutoTest (see Section 3.3.3).

The display is:

1psi 2MPa 3kPa

4bar

The unit of measure selection screen is the current [UNIT] selection screen (see
Section 3.3.2). Select the desired pressure measurement unit for the AutoTest.

The display is:

DUT full scale:

6000 psi g File1

Enter the full scale pressure of the DUT to be calibrated with this File AutoTest.
The value entered will be used along with the DUT tolerance to AutoRange
RPM4 settings when the File AutoTest is run (see Section 3.3.10,
 PRINCIPLE).

The display is:

DUT tolerance:

0.25% FS File1