4 commands, 1 programming messages, Commands – Fluke HPMS User Manual

Page 87: Programming messages, Table 14. program message list

Advertising
background image

4. REMOTE OPERATION

Page 77

© 2007 DH Instruments, a Fluke Company

4.4

COMMANDS

4.4.1

PROGRAMMING MESSAGES

Some commands accept an optional suffix. This suffix can be used to specify the explicit Q-RPT
to address. In most cases, each Q-RPT in the RPM4 operates as an independent measurement
device. If the suffix is not specified, then the “Active” Q-RPT is assumed.

The “RANGE” or “ARANGE” command can be used to select the “Active” Q-RPT. See the
program message descriptions for suffix use with a specific command.

Table 14. Program message list

COMMAND

n’ indicates an

optional suffix

DESCRIPTION

#

Send a command string out of the RPM4 COM2 port.

ABORT

Stop pending operations

ARANGEn

Read or set a new AutoRange to use.

ATM

Read the current atmospheric pressure (on-board barometer).

AUTOZEROn

Read or set the status of the AutoZ automatic zeroing function.

AUTOZEROn=RUN

Run AutoZero.

CALAMB

Read or set the on-board barometer calibration

COM1

Read or set the configuration of the COM1 port.

COM2

Read or set the configuration of the COM2 port.

CONTn

Enable continuous measurements

DATE

Read or set the current date.

ERR

Read the last error message.

GPIB

Read or set the GPIB interface address.

HEAD

Read or set the fluid head settings.

ID

Read or set the RPM4 alphanumeric asset ID tag.

L2

Selects “classic” program message format

L3

Selects “enhanced” program message format

LOCAL

Return control to the RPM4 front panel.

MEM

Read the power-up memory test status.

MMODEn

Read or change the active measurement mode.

MSGFMT

Read or set the type of program message format to use.

PCALn

Read or set the user Lo Q-RPT calibration information.

PCAL:XX

Read or set the user Lo Q-RPT calibration information (old command).

PRn

Read the next RPM4 pressure.

PRRn

Read the next RPM4 pressure, rate, and ATM.

QPRRn

Read the last RPM4 pressure, rate and ATM.

RANGE

Select a Q-RPT to be “active” in it’s default full scale range

RATEn

Read the next available rate of change of pressure.

READRATEn

Read or set the Q-RPT measurement read rate and mode

READYCKn

Read or set a flag that is cleared by a Not Ready condition.

REMOTE

Enable remote local lockout operation.

RESn

Read or set the pressure display resolution for the current transducer and range.

RESET

Reset the RPM4 to default user parameters.

RPTn

Read the available Q-RPT data.

SCRSAV

Read or set the front panel screen saver period.

SN

Read the serial number of the RPM4.

SRn

Read the next available pressure status (Ready/Not Ready).

SSn

Read or set the stability required for a Ready condition.

SS%n

Read or set the stability required for a Ready condition (% span/ s).

TIME

Read or set the current time of day.

UCOEFn

Convert a pressure in Pascal to pressure in the current units.

UDU

Read or set the user defined pressure unit.

ULn

Read or set the upper limit for the current range.

UNITn

Read or set the pressure unit of measure for the current range.

VER

Read the RPM4 software version.

ZOFFSETn

Read or set the AutoZero Z

offset

for the specified Q-RPT.

ZOFFSET:XX

Read or set the AutoZ Z

offset

for the specified Q-RPT (old command).