Table 6. security levels – Fluke HPMS User Manual

Page 74

Advertising
background image

RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL

© 2007 DH Instruments, a Fluke Company

Page 64

The security levels are structured to support typical levels of operation as shown
in Table 6. Specifically, the security levels prevent execution of the functions
accessed by the key strokes marked by “•”:

Table 6. Security levels

FUNCTION

LOW

MEDIUM

HIGH

[RANGE]

[ENT] (AutoTest)

[UNIT]

[MODE]

[AutoRange]

• •

[LEAK CK]

[DISPLAY]

• •

[HEAD]

[SDS] (NOT USED IN THIS RPM4 MODEL)

[AutoZ] (in absolute mode)

[SETUP]

[SETUP], <1range>

[SETUP], <2res>

• •

[SETUP], <4UL>

[SETUP], <5ATest>

[SPECIAL]

[SPECIAL], <1AutoZ>

[SPECIAL], <1AutoZ>, <1on/1off>

• • •

[SPECIAL], <1AutoZ>, <3edit>

[SPECIAL], <2remote>

[SPECIAL], <2remote>, make changes

[SPECIAL], <3head>

• •

[SPECIAL], <4SDS> (NOT USED IN THIS RPM4 MODEL)

[SPECIAL], <5pref>, <1ScrSvr>

• •

[SPECIAL], <5pref>, <2sound>

[SPECIAL], <5pref>, <3time>

• •

[SPECIAL], <5pref>, <3time>, make changes

SPECIAL], <5pref>, <4ID>, <2edit>

• • •

[SPECIAL], <6Punit>

[SPECIAL], <7internal>

• •

[SPECIAL], <7internal>, <3RPT2x> (NOT USED))

[SPECIAL], <7interrnal>, <5log>, clear log

• • •

[SPECIAL], <8cal>

[SPECIAL], <8cal>, <2edit> under any selection

• • •

[SPECIAL], <4reset>

[SPECIAL], <4reset>, <4cal>

• • •

[SPECIAL], <4reset>, <5all>

Remote communications disabled

“•” indicates the function/menu is NOT accessible.

Advertising
This manual is related to the following products: