The display is:

0.0 psi g Lo

Exercise DUT & [ENT]

The top line is the standard RPM4 run screen indicating the current measured
pressure (see Section 3.1.1). Exercise the DUT if desired using the top line indication
of the measured pressure and press [ENT] when complete.The display is:

1.

Standard RPM4 run screen indicating current
measured pressure.

2.

Number of this test point/the total number of
test points in the sequence.

3. Pressure to set for this increment in current

pressure unit of measure and measurement
mode.

0.0 psi g Lo

Set 0.0 1/11

Adjust the pressure to the <Set nnnn> value indicated. Set the pressure so
that the DUT, NOT the RPM4, indicates the <Set nnnn> value.

Once the specified pressure has been set, press [ENT]. If the RPM4 pressure
reading is in Not Ready condition (see Section 3.2.1), the RPM4 makes an
invalid entry sound and displays <Pressure NOT READY, TRY AGAIN>.
Correct the conditions that are causing the RPM4 reading to be excessively
unstable and press [ENT] again. If the RPM4 pressure reading is in Ready
condition, the RPM4 makes a valid entry sound and the display is:

1.

Standard RPM4 run screen indicating current
measured pressure.

2.

Number of this test point/the total number of
test points in the sequence.

3.

<F> to indicate freeze (see Section 3.3.6.7)
followed by the RPM4 pressure reading in
current units of measure captured when [ENT]
was pressed in the previous <Set> screen.
This is the reference pressure that was
applied to the DUT when the DUT indicated
the cardinal point. The frozen pressure value
flashes if the reading was out of tolerance
(definition of out of tolerance is: Set point –
RPM4 reading > DUT tolerance)..

497.7 psi g Lo

F497.6 OT 2/11

Press [ENT] to accept this test point and proceed to the next test point or [

] to

repeat the point.

Repeat the set pressure, take reading and accept data procedure until all points
in the sequence have been accepted.

After the last pressure point in the sequence
has been accepted the display is:

AutoTest complete

1data 2new 3repeat

Select <1data> to view the data collected in the AutoTest that was just
completed (see Section 3.4.5.1).

Select <2new> to run a new Quick or File AutoTest.

Select <3repeat> to repeat the AutoTest that was just completed.

Press [ESCAPE] to return to the main run screen.