2 as received and as left data, 2 equipment required, As received and as left data – Fluke HPMS User Manual

Page 118: Equipment required

Advertising
background image

RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL

© 2007 DH Instruments, a Fluke Company

Page 108

As editing PA and PM values changes Q-RPT calibration, they should only be

edited by qualified personnel as part of the calibration process. Caution should be
taken to avoid accidental editing and a security system is available to prevent access

(see Section 3.5.5.5). Incorrect editing of PA and PM values can cause out of

tolerance measurements.

A new RPM4 is delivered with PA and PM values set to zero and 1 for all

Q-RPTs. This does not mean that the RPM4 has not been calibrated. In the original

factory calibration, privileged factory coefficients are used for calibration with the

user PA and PM set to zero and 1.

5.3.1.2

AS RECEIVED AND AS LEFT DATA

Frequently, calibration procedures require that as received and as left data be
reported. The necessary information to report as received and as left data on the
calibration of RPM4 Q-RPTs can be obtained in several ways.

When the RPM4 CalTool calibration assistance software is used, as received
data is displayed while running the calibration and is automatically recorded and
reported if desired. As left data is also calculated and presented.

At any time, a) reference pressures applied; b) associated Q-RPT readings;
c) PA, PM and P

offset

(AutoZ offset) values, can be used to calculate as received and as

left values. For example, backing out PA and PM on the as left data yields the Q-
RPT readings with PA = 0 and PM = 1. Then applying the as received PA, PM
and P

offset

values to the readings calculates as received readings (the readings

that the transducer would have made with the old PA, PM and P

offset

).

It is recommended that “as received” values of PA, PM and P

offset

be recorded for

each range prior to running the calibration. The current PA, PM and P

offset

can be

viewed by pressing [SPECIAL], <8cal>, <1view>. P

offset

is only relevant if absolute mode

AutoZero is used in normal operation.

5.3.2

EQUIPMENT REQUIRED

The recommended calibration standard for the RPM4 A70M/A20M-AF Q-RPTs is a DHI

PG7202 piston gauge with a DHI RPM4 A100K reference pressure monitor connected to the

PG7202 as its external barometer. Contact DHI for additional information.

Gas operated piston gauge (deadweight tester), with the following characteristics:

Do not put oil or other liquids into RPM4 A70M/A20M-AF Q-RPTs.

Measurement uncertainty of ± 0.005 % of reading or better. A standard with higher

measurement uncertainty may be used but RPM4 measurement uncertainty may be
degraded proportionally from normal specifications.

Able to apply absolute pressures. If, after calibration, the RPM4 will be used in

absolute measurement mode, the calibration must be performed using absolute

Advertising
This manual is related to the following products: