RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL

© 2007 DH Instruments, a Fluke Company

Page 2

1.2

SPECIFICATIONS

1.2.1

GENERAL SPECIFICATIONS

Power Requirements:

85 to 264 VAC, 50/60 Hz, 25 VA max consumption
and 12VDC, 1.2 A

Operating Temperature Range:

18 to 28 °C

Storage Temperature Range:

- 20 to 70 °C

Weight:

11 kg (22 lb) approx.

Dimensions:

23.5 cm H x 29.4 cm W x 37.5 cm D (9.3 in. x 11.6 in. x 14.8 in.)

Communication Ports

Fuses:

RS232 (COM1, COM2), IEEE-488.2

1 A, 250 VAC fuse, 5 x 20 mm, time lag type fuse

Internal power supply fuse not replaceable by operator: 2.5A, 250 VAC

Pressure Range:

0 to 10 000 psi (70 MPa) in gauge mode

Atmosphere to 10 000 psi (70 MPa) in absolute mode

Operating Medium:

Any clean, dry, non-corrosive gas

Pressure Connections:

DH500 female

DH500 is a gland and collar type fitting for 6mm (1/4 in.) coned and left

hand threaded tubes equivalent to AE F250C, HIP HF4, etc.

Pressure Limits:

Maximum Working

Pressure: 10 000 psi (70 MPa)

Maximum Pressure Without Damage: 12 000 psi (83 MPa)

When low Q-RPT is connected to pressure:

Maximum Working

Pressure: 3 000 psi (20 MPa)

Maximum Pressure Without Damage: 3 600 psi (24 MPa)

PGC-10000-AF SYSTEM

Temperature:

Operating:

18 to 28

°C

Storage:

-20 to 70

°C

Relative Humidity:

Operating:

15 to 70%RH (non-condensing)

Storage:

10 to 90%RH (non-condensing)