RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL

© 2007 DH Instruments, a Fluke Company

Page 12

USE THE CORRECT PRESSURE CONNECTORS: The RPM4/HPMS TEST port fitting is a

DH500 F (see Section 1.2.1). It is NOT a 1/8 in. NPT F. Never use a fitting other than the
corresponding male fitting in these connectors. Damage to the connectors and dangerous failure

under pressure could result from using incorrect fittings.

DO NOT APPLY PRESSURE UNTIL YOU ARE FAMILIAR WITH OPERATION: The RPM4/HPMS

rear panel test port connects internally to both the RPM4’s 3 000 psi (20 MPa) and 10 000 psi

(70 MPa) Q-RPTs. The valve on the front of the HPMS isolates the low pressure Q-RPT when the

high pressure Q-RPT is in use. Do not apply pressure to the RPM4/HPMS until you are familiar
with its operation and know how to protect the low pressure Q-RPT from overpressure (see

Section 3.2.5). FAILURE TO PROTECT THE LOW PRESSURE Q-RPT FROM OVERPRESSURE MAY

DESTROY IT. DAMAGE DUE TO Q-RPT OVERPRESSURE IS NOT COVERED BY THE PRODUCT

WARRANTY.

2.4.6.1

THE ATM PORTS OF RPM4

The ATM ports on the RPM4 Q-RPT modules are connected to the RPM4’s
internal barometer to assure that the RPM4 gauge pressure measurements are
relative to ambient pressure. These ports should always be left completely
unobstructed and open to atmosphere.

NEVER plug, obstruct or connect a supply pressure to the RPM4 Q-RPT module

ATM ports. This may adversely affect GAUGE mode operation and AutoZeroing
functions.

2.4.7

CHECK/SET SECURITY LEVEL

RPM4 has a security system based on user levels. By default, the security system is set to
“low”, which includes access restriction to internal calibration coefficients, and there is no
password required to change the security level. See Section 3.5.5.5 for information on the
security level system. As part of the RPM4 startup, determine the security level that is
appropriate for the RPM4 and set a password if desired.

RPM4 is delivered with the security level set to “low” to avoid inadvertent altering of critical

internal settings but with access to changing security levels unrestricted. It is recommended that the

low security level be maintained at all times and password protection be implemented if control

over setting of security levels is desired.

2.4.8

SETTING UP AUTOTEST FILES

RPM4 supports automated test/calibration sequences. AutoTest sequence
parameters for testing specific DUTs can be stored in File AutoTest files and
recalled to run a test. Consider setting up File AutoTest files for frequently tested
DUTs as part of the RPM4 set up process (see Section 3.3.10, 3.4.5.2).