RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL

© 2007 DH Instruments, a Fluke Company

Page 24

3.2.6

DIRECT FUNCTION KEYS SUMMARY

Table 3 provides a brief summary of RPM4 direct function key operation. It may be useful to

keep a copy of this summary near the RPM4/HPMS, especially when first becoming acquainted with its

operation.

Local operation of RPM4 is through the front panel keypad. To minimize multi-layered menu
structures, the keypad numerical keys also provide direct access to the most commonly used
functions. The function accessed is labeled on the bottom half of the key. Direct function
keys are active whenever RPM4 is in its MAIN RUN screen. Table 3 summarizes the
operation of the direct function keys.

Table 3. Summary of RPM4 function key operation

DIRECT FUNCTION KEYS ARE ACTIVE FROM THE MAIN RUN SCREEN

See corresponding manual Sections for full detail.

Adjust height of fluid head
correction. Set to zero to defeat
correction.

View active range and toggle
through available ranges. [ENT] on
a range activates it.

The SDS function is not used in
the A70M/A20M-AF configuration
of RPM4.

Select pressure unit of measure for
the active range. Choice of units
available in this menu can be
customized.

Run the AutoZ function to
AutoZero the active Q-RPT and
measurement mode.

Select the pressure measurement
mode for the active range (absolute,
or gauge).

Set up and optimize RPM4
features for a specific operating
range and measurement mode.

Menu of less commonly used internal
functions and settings.

Run automated leak checking
routine.

Menu of commonly used setup
features including save/delete
AutoRange ranges, set display
resolution, set stability limit, set
upper/lower, set up File AutoTest
sequences and view AutoTest data.

Select the display function for the
bottom line of the RPM4 display.

Start or set functions such as Leak
Check, AutoTest and Freeze.

ENTER values when editing.

3.3

DIRECT FUNCTION KEYS

3.3.1

[RANGE]

 PURPOSE

To view and/or change the active pressure measurement range and associated settings.

 PRINCIPLE

An RPM4 range is defined as a pressure span and associated settings (see Section 3.2.3).

An RPM4 may have multiple ranges. Each Q-RPT in the RPM4 system has a default range
which is its maximum span. The default ranges of the RPM4 A70M/A20M are 10 000 psi
(A70M) and 3 000 psi (A20M). Additional ranges, lower than the Q-RPT’s maximum span,
may also be created using AutoRange (see Section 3.3.4) or AutoTest (see Section 3.3.10).