4 rpt – Fluke HPMS User Manual

Page 45

Advertising
background image

3. OPERATION

Page 35

© 2007 DH Instruments, a Fluke Company

With the Deviation DISPLAY active the main run screen is:

1.

Deviation of current pressure from the target
value.

2. Target

value.

8998.2 psi a zH3

D -1.8 T9000.0

Pressing [ENT] from the main run screen when the Deviation DISPLAY is active

goes directly to the Target editing screen. This allows the target value to be changed

and without going through the DISPLAY menu.

If the pressure measurement unit or mode is changed while the Deviation

DISPLAY is active the target value remains at the same numerical value. It is NOT

converted.

To go to a DISPLAY other than Deviation, press [DISPLAY] and make a new

DISPLAY choice (see Section 3.3.6).

3.3.6.4

RPT

 PURPOSE

To activate the RPT DISPLAY.

See Section 3.3.6,  Principle.

 OPERATION

For the sake of clarity, when describing the RPT DISPLAY, the Q-RPT displayed on

the top line of the display when the RPT function is selected is referred to as the “active”

Q-RPT. The other Q-RPT to be displayed on the second (bottom) line of the RPM4 display

is referred to as the “inactive” Q-RPT. It is “inactive” in the sense that all RPM4 functions

and settings such as UNIT and RES still apply to the “active” Q-RPT. To make changes to

the “inactive” Q-RPT, for example to change its pressure measurement unit, it must be
made the active Q-RPT by a conventional range change (see Section 3.3.1).

To activate the RPT DISPLAY, press [DISPLAY], <4RPT>. Operation returns to
the MAIN RUN screen with the inactive Q-RPT displayed on the display’s bottom
line. The range of the inactive Q-RPT is always the default range, not an
AutoRanged range (see Section 3.2.3).