3. OPERATION

Page 51

© 2007 DH Instruments, a Fluke Company

The over pressure function causes the measured pressure display to flash. The
overpressure function also logs the time and date of the overpressure condition in
both user and factory logs to assist in incident diagnosis (see Section 3.5.7.5).

To recover from an overpressure condition, cycle RPM4 power. Before cycling
power, be sure to correct the condition that caused the overpressure.

3.4.5

<5AT

EST

>

 PURPOSE

To view data collected while running an AutoTest and/or to view and edit File AutoTest
definition files.

See Section 3.3.10, PRINCIPLE for information on the AutoTest function.

 OPERATION

To access the ATest menu select [SETUP], <5ATest>. The display is:

Select <1data> to view AutoTest results data (see
Section 3.4.5.1).

Select <2file> to view or edit an AutoTest definition
file (see Section 3.4.5.2).

AutoTest:

1data 2file

3.4.5.1

DATA

 PURPOSE

To view the data files that are created when an AutoTest is run.

 PRINCIPLE

RPM4 supports AutoTest automated tests (see Section 3.3.10). When an
AutoTest is run, the RPM4 pressure measurement accepted by the operator at
each point is logged. The logged pressure measurements and an identifying
header are stored in a data file. The data file can be viewed immediately
following the last point of the sequence by selecting <1data>. In addition, the
last ten data files recorded can be viewed anytime by selecting [SETUP],
<5ATest>, <1data>.

Data File Protocol

Up to ten data files are buffered, ten at a time, in RPM4 memory. When a new
AutoTest run is completed, its data file goes to the back of the queue and the
oldest data file is deleted.

Data files are identified by a header whose first line is the full scale of the DUT
and the number of test points in the AutoTest. The second line is the date
(YYYYMMDD) and time (HH:MM:SS) at which the last point of the sequence
was accepted.

Following the data file header, each test point is recorded including the pressure
reading of the RPM4 with head, AutoZ and range status as well as an indication
of whether the reading was out of tolerance and the test point number over the
total number of test points in the Sequence.

 OPERATION

To access the AutoTest results data file viewing function press [SETUP],
<5ATest>, <1data>.

The first display is the data file header which can be used to toggle through
data files: