Use the “UP” or “DOWN” buttons to change the number from 2 to 8.

Pn=3

New number of program set-points

Press “SET” to continue. Pressing “EXIT” will cause any changes made to the
parameter to be ignored.

S

Save new setting

8.6.2

Set-points

The next parameters are the program set-points.

1 25.0

First set-point

Use the “UP” or “DOWN” buttons to select any of the set-points.

3 10.0

Third set-point

Press “SET” to change the set-point.

C 10.00

Set-point value

Use “UP” and “DOWN” to change the set-point value.

C 13.50

New set-point value

Press “SET” to save the new set-point value.

The other set-points can also be set in the same manner. Once the set-points are
programmed as desired press “EXIT” to continue.

E

Continue to next menu function

8.6.3

Program Soak Time

The next parameter in the program menu is the soak time. This is the time, in
minutes, for which each of the program set-points will be maintained after set-
tling before proceeding to the next set-point. The duration is counted from the
time the temperature settles to within a specified stability. The stability require-
ment can be set in the parameter menu as explained in Section 8.11.2.3. The
default is 0.1°C.

Pt=15

Soak time in minutes

Use the “UP” or “DOWN” buttons to change the time.

Pt=5

New soak time