The parameter is indicated by,

CtorSt

Cut-out reset mode parameter

Press “SET” to access the parameter setting. Normally the cut-out is set for au-
tomatic mode.

Cto=Auto

Cut-out set for automatic reset

To change to manual reset mode press “UP” and then “SET”.

Cto=rSt

Cut-out set for manual reset

8.11.2.3

Soak Stability

The soak stability controls the required stability of the well temperature for the
soak time (Section 8.6.3). The stability is in degrees Celsius. The default is
0.1°C. This value can be changed in the parameter menu.

8.11.3

Serial Interface Parameters

The serial RS-232 interface parameters menu is indicated by,

SErIAL

Serial RS-232 interface parameters menu

The serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to instruments fitted
with the serial interface. The parameters in the menu are — BAUD rate, sample
period, duplex mode, and linefeed.

8.11.3.1

BAUD Rate

The BAUD rate is the first parameter in the menu. The BAUD rate setting de-
termines the serial communications transmission rate.

The BAUD rate parameter is indicated by,

BAUd

Serial BAUD rate parameter

Press “SET” to choose to set the BAUD rate. The current BAUD rate value will
then be displayed.

1200 b

Current BAUD rate

The BAUD rate of the serial communications may be programmed to 300, 600,
1200, or 2400 BAUD. Use “UP” or “DOWN” to change the BAUD rate value.

2400 b

New BAUD rate

9105/9107

39