Marathon monitors inc – Marathon Computer AACC 2000 User Manual

Page 9

Advertising
background image

Marathon Monitors Inc.

AACC 2000 Carbon Nov. 1, 1997

9

General

Main PV Input range:

+100mV, 0 to 10Vdc (auto ranging) and 3 wire Pt100.

Calibration accuracy:

The greater of +0.2% of reading, +1 LSD or +1

o

C.

Cold junction compensation >30:1 rejection of ambient temperature, (for thermocouple

i/p).

Electrical safety

Standards:

EN 61010, Installation category II, pollution degree

2.
CSA C22.2 No.142-M1987.

Installation category II:

Voltage transients on any mains power connected to the
instrument must not exceed 2.5kV.

Pollution degree 2:

Conductive pollution must be excluded from the

cabinet in which the instrument is mounted.

Isolation:

All isolated inputs and outputs have reinforced

insulation to provide protection against electric shock. The
fixed digital inputs, non-isolated dc, logic, and the logic output
of dual output modules, are all electrically connected to the
main process variable input, (thermocouple etc.).