SECTION 5—CENTER MOUNT FOOTRESTS

Part No 1143238

19

Adjustable ASBA Seat

Adjusting the Height of the Manual Center Mount
Footrest

Chairs built before 12/01/2008

ƽ WARNING

After adjusting the center mount footrest, minimum ground clearance for the
footrest in the driving position is three inches. If the center mount footrest is in the
retracted position and the wheelchair is not moving, the footrest must maintain a
minimum of one inch ground clearance - otherwise personal injury and damage may
result.

NOTE: For this procedure, refer to FIGURE 5.2.

1. Remove the two mounting screws that secure the footrest extension tube to the 

extension tube housing.

2. Adjust the footrest extension tube to the desired height.

3. Align the mounting holes on the extension tube housing with the mounting holes on 

the footrest extension tube.

4. Reinstall the two mounting screws to secure the footrest extension tube to the 

extension tube housing. Securely tighten.

5. Repeat STEPS 1‐3 for the other extension tube.

FIGURE 5.2 Adjusting the Height of the Manual Center Mount Footrest -

Chairs built before 12/01/2008

Release Lever

Seat Frame

Extension Tube

Housing

Footrest

Extension Tube

Mounting Holes

Mounting Screws

Mounting Holes