CALIBRATION

Page: 51

5.3.2 CALIBRATION GAS DELIVERY METHODS

Calibration gas is can be delivered to the sensors via the
following delivery devices:

Model 5358-00: Calibration Adapter - used with portable
calibrators. - See Figure 5-1

Model 5360-00: Calibration Gas Delivery fitting -
permanently installed fitting which allows tubing to be run
to a convenient delivery location

5.3.3 SENSOR EXPOSURE TO GAS

Calibration gas must be delivered to the sensor using the
flow rate and duration listed in Table 5-1

Model

Gas

Flow

Period

5100-01

H

2

S

300 cc/min

3 minutes

5100-02

Combustible

100 cc/min

3 minutes

5100-03

O

2

200 cc/min

3 minutes

5100-04

CO

150 cc/min

3 minutes

5100-05

H

2

S

300 cc/min

3 minutes

5100-06

CL

2

300 cc/min

3 minutes

5100-07

H

2

300 cc/min

3 minutes

5100-10

SO

2

300 cc/min

3 minutes

5100-12

NO

2

300 cc/min

3 minutes

5100-13

CO

300 cc/min

3 minutes

5100-16

CO

300 cc/min

3 minutes

5100-19

NO

300 cc/min

3 minutes

5100-21

HCL

1,000 cc/min

10 minutes

5100-22

HCN

300 cc/min

3 minutes

5100-25

NH

3

300 cc/min

3 minutes

5100-26

HF

300 cc/min

3 minutes

5100-27

ETO

1,000 cc/min

3 minutes

Table 5-1

Calibration Gas Flow Rates

5.3.3.1 HYDROGEN SULFIDE

Hydrogen Sulfide gas must be supplied from a permeation
tube device or from Sierra Monitor Model 9200 series
calibrator.

5.3.3.2 COMBUSTIBLE

For compliance with factory Mutual (FMRC) Approval,
the Sierra Monitor Model 1200-26 is the only Approved
calibration gas delivery device. Apply 50% LEL methane
in air or the calibration concentration of the gas to be
monitored at 100 cc/min using Model 5358-00 or 5358-01
Calibration Adapter and Model 5360-00 Gas Delivery
Fitting. For calibration with a standard gas and cross
reference to a secondary gas, see Appendix D.

NOTE

Cross reference factors are not FMRC approved.

5.3.3.3 OXYGEN

The calibration procedure is different for Oxygen because
zero gas is not required and the sensor will be exposed
only to span gas. This is a single point calibration.

Either

: Deliver calibration gas (clean air - 20.9%

Oxygen through the magnetic head at 100
cc/min for one minute and then remove the air
and the calibration head.

Or

: If the local air is known to be clean leave

the magnetic head in place for one minute and
then remove.

5.3.3.4 TOXIC GAS

Toxic Gas for may be delivered from Sierra Monitor
Model 9200 series calibrator or from appropriate
permeation devices.

Figure 5-1

Model 5358-00 Calibration Adapter