SERVICE

Page: 73

Unscrew the old sensor from the conduit hub, screw in the
new sensor and connect the wiring harness to the
transmitter electronics. Replace the transmitter into the
enclosure, make signal adjustments as described above,
and replace the enclosure cover.

After the sensor is installed:

Update "new sensor" status in the change
module mode.

Allow one hour of stabilization of the new
sensor.

Make the electrical sensitivity adjustment as
described above.

Calibrate the sensor module.