Sentry Industries 5000 User Manual

Page 93

Advertising
background image

SMC

sierra monitor corporation

Sentry Instruction Manual - Version 6

APPENDIX A - SPECIFICATIONS

Page: 87

Nitric Oxide (5100-19)

Range:

Factory 20 PPM, Max 100 PPM

Operating Temperature 4

o

F to 122

o

F (-20

o

C to 50

o

C)

Relative Humidity

15 - 99% RH

Response:

<15 seconds to 90% indication

Calibration Frequency

90 Days, Recommended

Zero Drift:

Less than 2% per month

Typical Life:

3 years in normal service

Warranty:

1 year

Hydrogen Chloride (5100-21)

Range:

Factory 20 PPM, Max 100 PPM

Operating Temperature 4

o

F to 122

o

F (-20

o

C to 50

o

C)

Relative Humidity

15 - 99% RH

Response:

<100 seconds to 90% indication

Calibration Frequency

90 Days, Recommended

Zero Drift:

Less than 1% per month

Typical Life:

2 years in normal service

Warranty:

1 year

Hydrogen Cyanide (5100-22)

Range:

Factory 20 PPM, Max 200 PPM

Operating Temperature 4

o

F to 122

o

F (-20

o

C to 50

o

C)

Relative Humidity

15 - 99% RH

Response:

<70 seconds to 90% indication

Calibration Frequency

90 Days, Recommended

Zero Drift:

Less than 1% per month

Typical Life:

2 years in normal service

Warranty:

1 year

Ammonia (5100-25)

Range:

Factory 50 PPM, Max 50 PPM

Relative Humidity

20 - 90% RH

Operating Temperature 14

o

F to 113

o

F (-10

o

C to 45

o

C)

Response:

<30 seconds to 90% indication

Calibration Frequency

90 Days, Recommended

Zero Drift:

Less than 2% per month

Typical Life:

2 - 3 years in normal service

Warranty:

1 year

Ethylene Oxide (5100-27)

Range:

Factory 20 PPM, Max 100 PPM

Relative Humidity

5 - 99% RH

Operating Temperature 4

o

F to 122

o

F (-20

o

C to 50

o

C)

Response:

<90 seconds to 90% indication

Calibration Frequency

90 Days, Recommended

Zero Drift:

Less than 2% per month

Typical Life:

2 years in normal service

Warranty:

1 year

Model

Gas

Reading from 100 PPM of interfering gas

Number

Type

CO

H

2

S

Cl

2

H

2

SO

2

NO

2

NO

HCl

HCN

C

2

H

4

5100-04

CO

100

315

-15

<40

50

-55

30

2

40

<50

5100-05

H

2

S

<0.5

100

-20

<0.1

<15

-15

0

0

0

0

5100-06

Cl

2

0

<-10

100

0

0

105

0

0

0

0

5100-07

H

2

<1

<20

0

100

3

0

35

3

35

85

5100-10

SO

2

<1

0

-40

0

100

-100

0

0

15

0

5100-12

NO

2

0

-20

90

0

0

100

0

0

-3

0

5100-13

CO

100

350

-40

<60

65

-60

25

5

40

90

5100-16

CO

100

0

0

0

0

0

0

0

0

35

5100-19

NO

0

35

0

0

5

25

100

15

0

0

5100-20

HCl

0

75

-10

0

35

-2

0

100

-8

0

5100-22

HCN

<0.5

-50

0

160

-190

-5

30

100

<1

Ethanol

Toluene

Methyl-ethyl-ketone

Carbon Monoxide

5100-27

C

2

H

4

O

55

20

10

40

Table 8-1

Toxic Sensor Cross Sensitivity

Advertising