Sentry Industries 5000 User Manual

Page 86

Advertising
background image

SMC

sierra monitor corporation

Sentry Instruction Manual - Version 6

SERVICE

Page: 80

7. Apply span gas. Adjust the gain potentiometer

until TP10 = correct value as described above.
To increase voltage at TP10 turn the gain
potentiometer counter clockwise.

8. Remove system power and re-install the sensor

and transmitter in the enclosure. Restore power
and calibrate.