F72 – Fault Status

Use function F72 to display the fault status of the clock.

Clock PLL – Locked or unlocked

Clock Status – Locked or unlocked

Command Line

To display the status of the fault detectors, enter:

F72<CR>

F72 responds:

F72 CLOCK PLL LOCKED

CLOCK STATUS UNLOCKED

where:

F73 – Alarm Control / Status

Use function F73 to do the following:

See the state of an indicator (“Locked/Unlocked” or “OK/Fault”)

Enable or disable the alarm for each indicator

See the state of the Alarm Latch for each indicator

Clear the Alarm Latch for all indicators

Enable or disable blinking of the Alarm Status LED on the front panel while it is green or amber

Set the values for Time Threshold, Timeout Delay, and Power-On Alarm Suppress

The following table summarizes F73’s alarm indicators and parameters, as well as the factory settings
for an XL-GPS without options. The factory settings vary depending on the options included at the XL-
GPS at the time it ships from the factory.

CLOCK PLL

= Clock Phase Loop Lock status, LOCKED or UNLOCKED

CLOCK STATUS

= Clock Status, LOCKED or UNLOCKED to the reference source shown.

The LOCKED state is based on the F73 Time Threshold value entered
by the operator. F73 Time Threshold sets the threshold in nS above
which Time Error triggers an alarm. When the oscillator's predicted
worst-case time error (“F13 – Time Error” on page 48) exceeds F73
Time Threshold, Time Error enters a fault state and the CLOCK
STATUS becomes UNLOCKED.