10 gain and time constant menu – SRS Labs SR850 User Manual

Page 120

Advertising
background image

5-10

Gain and Time Constant Menu

Reserve

This key selects either maximum, minimum or manual dynamic reserve.
When the reserve is maximum, the SR850 automatically selects the max-
imum reserve available at the present full scale sensitivity. When the
reserve is minimum, the minimum available reserve is selected. The
dynamic reserve is displayed in the settings display at the top of the
screen (if selected in the DISPLAY/SCALE menu).

When the reserve is manual, the reserve is set by the user to a value
between the min and max reserve. The reserve is set in 10 dB
increments.

Changing the sensitivity may change the reserve. With max or min
reserve, the max or min reserve of the new sensitivity is used. With
manual reserve, the reserve is kept the same relative to the mini-
mum reserve.
For example, if the sensivity is 10 mV and the reserve is
21 dB (20 dB above the minimum for this sensivity), changing the sensiv-
ity to 5 mV will change the reserve to 27 dB (20 dB above the 7 dB mini-
mum). If the new sensivity does not have a reserve high enough to satis-
fy this rule, then the max sensitivity will be used instead.

The minimum and maximum reserve limits for each sensitivity are listed
below.

Sensitivity

Min Reserve (dB)

Max Reserve (dB)

1 V

0

0

500 mV

6

6

200 mV

4

14

100 mV

0

20

50 mV

6

26

20 mV

4

34

10 mV

0

40

5 mV

6

46

2 mV

4

54

1 mV

10

60

500 µV

16

66

200 µV

24

74

100 µV

30

80

50 µV

36

86

20 µV

44

94

10 µV

50

100

5 µV

56

106

2 µV

64

114

1 µV

70

120

500 nV

76

126

200 nV

84

134

100 nV

90

140

50 nV

96

146

20 nV

104

154

10 nV

110

160

5 nV

116

166

2 nV

124

174

Do not use ultra high dynamic reserves above 120 dB unless absolutely
necessary. It will be very likely that the noise floor of any interfering