32 remote programming – SRS Labs SR850 User Manual

Page 208

Advertising
background image

6-32

Remote Programming

LIA STATUS BYTE

bit

name

usage

0

INPUT/RESRV

Set when an INPUT or RESRV overload is
detected.

1

FILTR

Set when a FILTR overload is detected.

2

OUTPT

Set when an OUTPT overload is detected.

3

UNLK

Set when a reference unlock is detected.

4

RANGE

Set when the detection frequency switches
ranges (harmonic x ref. frequency decreases
below 199.21 Hz or increases above
203.12 Hz). Time constants above 30 s and
Synchronous filtering are turned off in the upper
frequency range.

5

TC

Set when the time constant is changed indirect-
ly, either by changing frequency range, dynamic
reserve or filter slope.

6

TRIG

Set when unit is triggered. Only if samples or
scans are triggered.

7

PLOT

Set when a plot is completed.

The LIA Status bits stay set until cleared by reading or by the

CLS command.

ERROR STATUS BYTE

bit

name

usage

0

Prn/Plt Error

Set when an error occurs during printing or
plotting.

1

Backup Error

Set at power up when the battery backup has
failed.

2

RAM Error

Set when the RAM Memory test finds an error.

3

Disk Error

Set when an error occurs during a disk or file
operation.

4

ROM Error

Set when the ROM Memory test finds an error.

5

GPIB Error

Set when GPIB fast data transfer mode aborted.

6

DSP Error

Set when the DSP test finds an error.

7

Math Error

Set when an internal math error occurs.

The Error Status bits stay set until cleared by reading or by the

CLS command.