background image

8-4

SR850 Service

If the R reading (top display) is less than 1 mV, then proceed to the next step. If the reading is
greater than 1 mV, adjust the Voltage Input Offset pot until R is less than 1 mV.

8. Press

<Sensitivity>

Rotate the knob to select 1 mV full scale

If the R reading is less than 0.02 mV then proceed to the next step. If the reading is greater than
0.02 mV, adjust the Voltage Input Offset pot until R is less than 0.02 mV.

9. Connect the SINE OUT output to both the A and B inputs. Use equal length cables from A and B

to a BNC TEE. Connect the cable from SINE OUT to the TEE.

Press

[GAIN/TC]

<Sensitivity>

Rotate the knob to select 1 V full scale

[REF/PHASE]

<Ref. Frequency> Select Reference Frequency

[1] [0] [0] [ENTER] Enter 100 Hz

[INPUT/FILTERS]

<Coupling> Select DC coupling

The R reading (top display) should read 1.00 V (±2%).

Press

<Source> to select A-B

If the R reading is less than 0.001 V, then proceed to the next step. If the reading is greater than
0.001 V, adjust the Common Mode pot until R is less than 0.001 V.

Press

[GAIN/TC]

<Sensitivity>

Rotate the knob to select 2 mV full scale

If the R reading is less than 0.2 mV, then proceed to the next step. If the reading is greater than
0.2 mV, adjust the Common Mode pot until R is less than 0.02 mV.

Press

[GAIN/TC]

<Sensitivity>

Rotate the knob to select 200 µV full scale

If the R reading is less than 2 µV, then proceed to the next step. If the reading is greater than
2 µV, adjust the Common Mode pot until R is less than 2 µV.

10. The offset needs to be readjusted after adjusting the common mode. Connect a 50

termination

to the A input.

Press

[INPUT/FILTERS]

<Source> twice to select A
<Coupling> to select AC

[REF/PHASE]

<Ref. Frequency> Select Reference Frequency

[1] [ENTER] Enter 1 Hz