Kodak i1800 Series User Manual

Page 42

Advertising
background image

3-26

A-61555 March 2007

UDDS Calibration

UDDS calibration is used to calibrate the ultrasonic document detection
system for multi-feeds. UDDS calibration is rarely needed. This
calibration should only be performed if there is a feeding or multi-feed
detection issue.

1. Touch Calibrate. The Calibration screen will be displayed.

2. Touch UDDS Calibration to display the UDDS Calibration screen.

3. You will be prompted to place a 20 lb. or 75 g/m

2

bond sheet of

paper in the input elevator. Do not use the calibration target.

4. Touch OK. The calibration results will be displayed.

5. When finished, touch End Job to return to the Idle screen.

Advertising