4 process current, 5 process voltage – Omega ILD44-UTP User Manual

Page 15

Advertising
background image

11

2.3.4 Process Current

The figure below shows the wiring hookup for Process Current 0 – 20 mA.

Figure 2.8 Process Current Wiring Hookup

(Internal and External Excitation)

When configuring your instrument, select Process Type in the Input Type Menu (see Part 3).

2.3.5 Process Voltage

The figure below shows the wiring hookup for Process Voltage 0 – 100 mV,
0 – 1 V, 0 – 10 V.

Figure 2.9
a) Process Voltage Wiring Hookupb) Process Voltage Wiring Hookup

with Sensor Excitation

without Sensor Excitation

R

L

- Voltage limiting resistor, which allows conversion of the 24 Vdc internal excitation voltage to

the appropriate process input value. For instance: if the potentiometer value is equal to 10 kΩ, the
minimum R

L

is 14 kΩ for 10 V process input.

When configuring your instrument, select Process Type in the Input Type Menu (see Part 3).