CE APPROVAL INFORMATION

1. Electromagnetic Compatibility (EMC)

This device conforms with requirements of EMC Directive 89/336/EEC,
amended by 93/68/EEC. This instrument complies with the following EMC
Immunity Standards as tested per EN 61326-2:1998 (Industrial Locations)

Phenomena

Test Specification

Basic Standard

Electrostatic

+/- 4 kV contact discharge

IEC 1000-4-2

Discharge

+/- 8 kV air discharge

Performance
Criteria B

Radio Frequency

27 - 1000 MHz

IEC 1000-4-3

electromagnetic

3 V/m

Performance

field.

80% AM (1 KHz)

Criteria A

Fast Transients

+/- 2 kV (ac mains)

IEC 1000-4-4

+/- 1 kV (dc, signal I/O)

Performance

5/50 ns Tr/Th, 5 KHz rep. freq.

Criteria B

Surge

+/- 1.0 kV AC Port L-L

IEC 1000-4-5

+/- 2.0 kV AC Port L-PE

Performance

+/- 1.0 kV DC Port L-L

Criteria B

+/- 2.0 kV DC Port L-PE

Radio Frequency

0.15 - 80 MHz

IEC 1000-4-6

conducted

3 Vrms

Performance

80% AM (1 KHz)

Criteria A

This instrument complies with the following EMC Emission Standards as
tested per EN 61326-2:1998 (Class B equipment)

Phenomena

Frequency

Limits

Basic

Range

Standard

Radiated

30-230 MHz

30 dB_V/m at 10 m

CISPR 22

Emission

230-1000 MHz

37 dB_V/m at 10 m

Class B

quasi peak

Conducted

0.15-0.5 MHz

66-56 dB_V quasi peak

CISPR 22

Emission

0.5-5 MHz

56 dB_V quasi peak

Class B

5-30 MHz

60 dB_V quasi peak

2.Safety

This device conforms with Low Voltage Directive 73/23/EEC, amended by 93/68/EEC.
The following LVD requirements have been met to comply with EN 61010-1, 1993
(Electrical equipment for measurement, control and laboratory use)

1. Pollution Degree 2
2. Installation Category II
3. Class I Equipment (100-240 Vac Powered Units)