Tyco 1540 User Manual

Page 49

Advertising
background image

4.2.3 Measuring Ranges

Values of measured quantities for which errors are defined.

Voltage

70 .. 120% of nominal

Current

5 .. 120% of nominal

Frequency

45 .. 66 Hz, 360 .. 440 Hz (Model 0440)
Crest values of voltage and current must remain within
168% of nominal maximum rms values

4.2.4 Accuracy

Voltage

0.4% of reading ±0.1% of range
1% of range maximum for Model 0440

Current

0.4% of reading ±0.1% of range
1% of range maximum for Model 0440

Frequency (not 0340)

0.15% of mid frequency
1% of mid frequency for Model 0440

Temperature coefficient

0.013%/°C typical

Response time to step input

1.5 seconds approx.

Screen update time

0.5 second approx.

4.2.5 Reference conditions of influence quantities

Values that quantities which affect measurement errors to a minor degree

have to be for the intrinsic (headline) errors for measured quantities to apply.

Ambient temperature

23°C

Input frequency

50 or 60 Hz 2%

Input waveform Sinusoidal

(distortion factor 0.005)

Auxiliary supply voltage Nominal

1%

Auxiliary supply frequency Nominal

1%

Auxiliary supply distortion factor

0.05

Magnetic field of external origin

Terrestrial flux

4.2.6 EMC Standards

EMC Immunity EN61326 for Industrial Locations to performance criterion A

EMC Emissions EN61326 to Class B - Domestic

47

Integra 1540, 1000, 0640, 0440, 0340, 0240 Issue 1 04/03