Opc -31 *opt? -31 – Tektronix AFG3000 User Manual

Page 105

Advertising
background image

Syntax and Commands

AFG3000 Series Arbitrary/Function Generators Reference Manual

3-31

*OPC

This command generates the operation complete message by setting bit 0 in the
Standard Event Status Register (SESR) when all pending commands that generate
an OPC message are complete.

The query command places the ASCII character “1” into the output queue when all
such OPC commands are complete.

Group

Synchronization

Syntax

*OPC

*OPC?

Arguments

None

Returns

<execution complete>::=1
where “1” indicates that all pending operations are complete.

Examples

*OPC?
might return 1 to indicate that all pending OPC operations are finished.

*OPT?

This query-only command returns a list of the options installed in your arbi-
trary/function generator.

Group

System

Syntax

*OPT?

Arguments

None

Returns

<OPT>[,<OPT>[,<OPT>[,<OPT>]]]

Examples

*OPT?
might return 0, which indicates no option is installed in the instrument.

Advertising