Syntax and Commands

AFG3000 Series Arbitrary/Function Generators Reference Manual

3-35

*PSC

This command sets and queries the power-on status flag that controls the automatic
power-on execution of SRER and ESER. When *PSC is true, SRER and ESER are
set to 0 at power-on. When *PSC is false, the current values in the SRER and ESER
are preserved in nonvolatile memory when power is shut off and are restored at
power-on.

Group

Status

Syntax

*PSC <NR1>

*PSC?

Arguments

<NR1>=0 sets the power-on status clear flag to false, disables the power-on clear,
and allows the instrument to possibly assert SRQ after power-on.

<NR1>

≠0 sets the power-on status clear flag true. Sending *PSC 1 therefore

enables the power-on status clear and prevents any SRQ assertion after power-on.

Returns

<NR1>

Examples

*PSC 0
sets the power-on status clear flag to false.

*RCL (No Query Form)

This command restores the state of the instrument from a copy of the settings stored
in the setup memory. The settings are stored using the *SAV command. If the spec-
ified setup memory is deleted, this command causes an error.

Group

Memory

Related Commands

*SAV

Syntax

*RCL {0|1|2|3|4}

Arguments

0, 1, 2, 3, or 4 specifies the location of setup memory.

Examples

*RCL 3
restores the instrument from a copy of the settings stored in memory location 3.