System:klock[:state, System:security:immediate (no query form) – Tektronix AFG3000 User Manual

Page 170

Advertising
background image

Syntax and Commands

3-96

AFG3000 Series Arbitrary/Function Generators Reference Manual

SYSTem:KLOCk[:STATe]

This command locks or unlocks the instrument front panel controls. The query
command returns “0” (OFF) or “1” (ON).

Group

System

Syntax

SYSTem:KLOCk[:STATe] {ON|OFF|<NR1>}

SYSTem:KLOCk[:STATe]?

Arguments

ON or <NR1>

≠0 locks front panel controls.

OFF or <NR1>=0 unlocks front panel controls.

Returns

<NR1>

Examples

SYSTem:KLOCk:STATe ON
locks front panel controls.

SYSTem:SECurity:IMMediate (No Query Form)

This command erases all the current instrument setups, setup memory, last setup
memory, user waveform memory, and log content, and recalls the factory default
settings. Calibration data is not erased.

The communication settings are initialized to the factory default settings. This
might cause a remote communication error.

Group

System

Syntax

SYSTem:SECurity:IMMediate

Arguments

None

Examples

SYSTem:SECurity:IMMediate
initializes the instrument.