Source[1|2]]:pulse:dcycle, Source[1|2]]:pulse:dcycle -67 – Tektronix AFG3000 User Manual

Page 141

Advertising
background image

Syntax and Commands

AFG3000 Series Arbitrary/Function Generators Reference Manual

3-67

[SOURce[1|2]]:PULSe:DCYCle

This command sets or queries the duty cycle of the pulse waveform for the speci-
fied channel. The setting range is 0.1% to 99.9% in increments of 0.1.

The arbitrary/function generator will hold the settings of leading edge and trailing
edge when the duty cycle is varied.

Refer to the [SOURce[1|2]]:PULSe:WIDTh command for the setting range.

Group

Source

Related Commands

[SOURce[1|2]]:PULSe:WIDTh

Syntax

[SOURce[1|2]]:PULSe:DCYCle {<percent>|MINimum|MAXimum}

[SOURce[1|2]]:PULSe:DCYCle? {MINimum|MAXimum}

Arguments

<percent>::=<NR2>[<units>]
where
<NRf> is the duty cycle.
<units>::=PCT

Returns

<percent>

Examples

SOURce1:PULSe:DCYCle 80.5
sets the duty cycle of the pulse waveform on CH 1 to 80.5%.