Source[1|2]]:voltage[:level][:immediate]:low, Source[1|2]]:voltage[:level][:immediate]:low -83 – Tektronix AFG3000 User Manual

Page 157

Advertising
background image

Syntax and Commands

AFG3000 Series Arbitrary/Function Generators Reference Manual

3-83

[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:LOW

This command sets or queries the low level of output amplitude for the specified
channel. If your instrument is a dual-channel model and the
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe] command is set to ON, then the
low level of other channel is also the same value.

Group

Source

Related Commands

[SOURce[1|2]]:VOLTage:CONCurrent[:STATe]

Syntax

[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:LOW {<voltage>|MINimum|
MAXimum}

[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:LOW? {MINimum|MAXimum}

Arguments

<voltage>::=<NRf>[<units>]
where
<NRf> is the low level of output amplitude.
<units>::=[mV | V]

Returns

<voltage>

Examples

SOURce1:VOLTage:LEVel:IMMediate:LOW -1V
sets the low level of CH 1 output amplitude to –1 V.

Advertising